HZKS-3 Automatic COC Cleveland Open Cup Flash Point Apparatus
Function Overview
Flash Point is the lowest temperature at which a liquid can form an ignitable mixture in air near the surface of the liquid. The lower the flash point, the easier it is to ignite the material. Nowadays, Flash point is used to say non-volatile or flammable substances whether exit highly volatile or flammable substances, and has been one of the most important parameters for petroleum products transportation, storage, operations, safety management.
Technical Parameters
Temperature Measurement | |
Temperature measuring range |
-59.9°C~399.9°C |
Temperature Control Range |
Room temperature~400°C |
Repeatability |
≤150℃ 3℃, >150℃ 5℃ |
Resolution |
0.1°C |
Precision |
0.5% |
Temperature Sensor |
Platinum Resistor(PT100) |
Flash Fire Sensor |
Ion detection ring |
Ambient Temperature |
0 ~ 40°C |
Relative Humidity |
≤ 85% RH |
Power Supply |
AC220V ± 20%; 50HZ ± 10% |
Power Consumption |
Approx 400W |
Overall Dimension |
440X280X290(mm) |
Instrument Weight |
14kg |
Failure And Treatment
NO. |
Failure |
Treatment |
1 |
Turn on the power, the LCD screen no display |
Check power supply is plugged in, open the instrument measured the board examine the plug is not loose. |
2 |
Large repeatability error for sample testing |
Remove the lifting arm top cover, and check |
3 |
No heating |
Electric wire broken |
4 |
Lift arm rise or fall failure |
Limit opt coupler damage |
http://www.transformer-tester.com/